ANNEX 7 OF THE OMCL NETWORK GUIDELINE “QUALIFICATION OF EQUIPMENT” QUALIFICATION OF MASS SPECTROMETERS
Note: Mandatory requirements in this guideline and its annexes are defined using the terms «shall» or «must». The use of «should» indicates a recommendation. For these parts of the text other appropriately justified approaches are acceptable. The term «can» indicates a possibility or an example with non-binding character.
Introduction
The present document is the 7th Annex of the core document “Qualification of Equipment” and it shall be used in combination with it when planning, performing and documenting the qualification process of Mass Spectrometers coupled with Chromatographic equipment.
The core document “Qualification of Equipment” contains the introduction and general forms for Level I and II of qualification, which are common to all type of instruments.
The present Annex 7 contains a general introduction and requirements for GC-EI-MS and LC-ESI-MS.
Level III and IV qualifications must be carried out being an ISO 17025 requirement.
Requirements and (if applicable) corresponding typical acceptance limits given in bold should be applied, however other appropriately justified deviations are acceptable provided they are traceable.
Exemplary procedures provided in Annexes have non-binding character. They can be helpful to carry out the required qualification. Nevertheless, other procedures can be applied depending on the model of the MS equipment.
7.1: QUALIFICATION OF GAS CHROMATOGRAPHY-MASS SPECTROMETERS WITH ELECTRON IMPACT IONIZATION (GC-EI-MS)
TABLE I
Requirements and related typical acceptance limits are indicated in bold.
Parameter to be checked | Typical acceptance limits* |
Mass accuracy (PFTBA ** (FC-43) or internal calibration gas) |
m/z = 69.0 ±
0.5
m/z = 219.0 ±
0.5 |
Linearity*** | Limit to be set up based on OMCL experience/service provider instructions and type of regression mode chosen |
System/instrument precision*** | RSD ≤ 10.0 % |
*Other figures given under column of “typical acceptance limits” are typical values obtained when applying the exemplary procedures provided in annexes, therefore these values are not binding.
** PFTBA (FC-43): Perfluoro-tributyl-amine (CAS NO.: 311-89-7).
*** To be checked when quantification is requested.
TABLE II
Requirements and related typical acceptance limits are indicated in bold.
Parameter to be checked/ Typical acceptance limits |
According to specific analysis method or Ph. Eur. monograph or MAH dossier (see examples in Level IV, Annex II) |
ANNEX I
This Annex contains practical exemplary procedures for several parameters related to the qualification of GC-EI-MS.
These examples can be considered by the OMCLs as possible approaches to perform the Level III of the equipment qualification process: “Periodic and motivated instrument checks”.
However, alternative procedures can be applied.
GENERAL CONSIDERATION: GC-MS is mainly used for the identification of unknown substances or quantification of low concentrated substances where high specificity is needed.
MASS ACCURACY
Materials:
PFTBA (FC-43) or internal calibration gas
Method:
Internal instrument check or spectrum of PFTBA (FC-43) in full scan mode
Limits:
As defined in Table I
LINEARITY
Materials:
Stock solutions: 1-Octanol in methanol at concentrations of 0.2, 0.4, 0.6, 0.8, 1.0 µL/mL
Method:
Injection volume: 1.0 µL (2 injections of each level)
Limits:
Limit to be set up by OMCL based on experience and type of regression mode chosen
NOTE: Linearity is typically performed in SIM/MRM, since this mode is normally applied for quantification of analytes in low concentration.
SYSTEM/INSTRUMENT PRECISION
Materials:
Stock solution: 1-Octanol in methanol at concentration of 1.0 µL/mL
Method:
Injection volume: 1.0 µL (6 injections)
Limits:
RSD ≤ 10.0 % on peak area (without internal standard)
ANNEX II
IDENTIFICATION (by mass spectral library)
Materials:
Papaverine 20.0 µg/mL in methanol, Caffeine 10.0 μg/mL in methanol or another compound chosen according to the specific method
Method:
Identification by mass spectral library
Limits:
Match reference spectra
SYSTEM/INSTRUMENT PRECISION **
Materials:
See identification
Method:
Injection volume: 1.0 µL (6 injections)
Limits:
RSD ≤ 10.0 % on peak area
** to be checked when performing quantification testing
7.2 QUALIFICATION OF LIQUID CHROMATOGRAPHY-MASS SPECTROMETERS (LC-MS)
Ionisation sources of Mass Spectrometry include:
ESI: Electrospray Ionisation
APCI: Atmospheric Pressure Chemical Ionisation
APPI: Atmospheric Pressure Photo-Ionisation
TABLE III
Requirements and related typical acceptance limits are indicated in bold.
IDENTIFICATION - Positive mode:
Parameter to be checked | Typical acceptance limits (Low Resolution MS)* | Typical acceptance limits (High Resolution MS)* |
Mass accuracy (Reserpine) | m/z = 609.3 ± 0.5 | ± 5.0 ppm |
Mass accuracy of fragments** (Reserpine) |
m/z = 448.2 ±
0.5 m/z = 195.1 ± 0.5 |
±
5.0 ppm ± 5.0 ppm |
Resolution | See Annex III | See Annex III |
** only for instruments with MS/MS capabilities.
IDENTIFICATION - Negative mode:
Parameter to be checked | Typical acceptance limits (Low Resolution MS)* | Typical acceptance limits (High Resolution MS)* |
Mass accuracy (Chloramphenicol**) | m/z = 321.0 ± 0.5 | ± 5.0 ppm |
Mass accuracy of fragments *** (Chloramphenicol) | m/z = 152.0 ± 0.5 | ± 5.0 ppm |
Resolution | See Annex III | See Annex III |
* Other figures given under column of “typical acceptance limits” are typical values obtained when applying the exemplary procedures provided in annexes, therefore these values are not binding.
** Chloramphenicol, CAS NO.: 56-75-7.
*** only for instruments with MS/MS capabilities.
QUANTIFICATION:
Check following parameters both in positive and negative modes:
Parameter to be checked | Typical acceptance limits |
Linearity | Limit to be set up based on OMCL experience/service provider instructions and type of regression mode chosen |
System/instrument precision | RSD ≤ 5.0 % |
Carry over: | ≤ 1.0 % |
TABLE IV
Level IV. In-use instrument checks
Requirements for LC-ESI-MS
Parameter to be checked/ Typical acceptance limits |
According to specific analysis method or Ph. Eur. monograph or MAH dossier |
ANNEX III
This Annex contains practical exemplary procedures for several parameters related to the performance of LC-ESI-MS.
These examples can be considered by the OMCLs as possible approaches to perform the Level III of the equipment qualification process: “Periodic and motivated instrument checks”.
However, alternative procedures can be applied.
GENERAL CONSIDERATION: LC-MS is widely used for the identification of unknown substances or quantification of low concentrated substances where high specificity is needed.
MASS ACCURACY
Materials:
ESI positive: Reserpine at concentration of 0.01 mg/mL in methanol/water (60/40 V/V)*
ESI negative: Chloramphenicol at concentration of 0.01 mg/mL in methanol containing 0.1% of formic acid*.
*Note: concentrations should be selected depending on the instrument and experimental conditions to be applied
Method:
Direct infusion or flow injection
Limits:
As defined in Table III.
RESOLUTION:
Procedure to be followed depends on the instrument. Instructions to check resolution can be provided by service supplier/ instrument manual.
LINEARITY
Materials:
Solutions as mixture of betamethasone-17,21-dipropionate and betamethasone-17-valerate in methanol at concentrations of 0.002, 0.004, 0.006, 0.008, 0.01 mg/mL
Method:
Column: BEH-C18 1.7 µm 50 x 2.1 mm or equivalent
Suitable gradient of acetonitrile/water containing 0.1 % formic acid
Injection volume: 1.0 µL (2 injections at each concentration)
Limits:
Limit to be set up based on OMCL experience/service provider specifications and type of regression mode chosen
SYSTEM/INSTRUMENT PRECISION
Materials:
Solution as mixture of betamethasone-17, 21-dipropionate and betamethasone-17-valerate in methanol at concentration of 0.006 mg/mL.
Method:
Injection volume: 1.0 µL (6 injections)
Limits:
RSD ≤ 5.0 % on peak area
CARRY OVER
Materials:
- Solution as mixture of betamethasone 17, 21-dipropionate and betamethasone 17- valerate in methanol at concentration of 0.002 mg/mL.
- Methanol (blank)
Method:
Injection volume: 1.0 µL
Limits:
The percentage of the peaks corresponding to betamethasone-17,21-dipropionate and betamethasone 17-valerate in the blank (injected after 0.002 mg/mL solution) does not exceed 1.0 % of the area of said peaks in the chromatogram obtained injecting 0.002 mg/mL solution of betamethasone-17,21-dipropionate and betamethasone 17-valerate.
REFERENCES
(For all references, the latest version applies)
1) Ph. Eur. Chapter 2.2.43. MASS SPECTROMETRY.